5-Channel Qualification Kit Result Review Table
| Error | Acceptable Criteria | Initial Field Service Action | Follow-Up Run | Final Field Service Action |
|---|---|---|---|---|
| Quad Point Melt (Required to Run before EBP) | ||||
|
[cartridge row][channel] EVALUATE
Description: PCR heater exceeded set tolerances [±2] (530nm/565nm OR 585nm/630nm OR 630nm/665nm) OR [±3] (475nm/520nm) *Note optic 680/715nm data is not evaluated as a part of the QPM |
0 Evaluate results |
|
Only rerun specific bank of suspect lane | Replace heater/MUX board ONLY if same lane triggers error |
| EBP Qualification | ||||
|
No Amp on > 2 Lanes WITHIN Rack and Row; on Color [color] [cartridge row] Description: Error triggered only if >2 lanes on the same reader and channel
|
0 Evaluate results |
In addition, perform the following:
|
Only rerun specific bank of suspect reader | Replace reader ONLY if same reader triggers error |
|
IC No Amp [cartridge row]
Description: Error triggered only if in channel Cy5.5 (680nm/715nm) If "No Amp on >2 Lanes" error also raised, overlook error |
0 Evaluate results |
In addition, perform the following:
|
Only rerun specific bank of suspect lane | Escalate through your Regional Escalation Process. |
|
Target NO Amp on Color [color][cartridge row]
Description: Error triggered only if in channels 530nm/565nm OR 585nm/630nm OR 630nm/665nm OR 475nm/520nm
If "No Amp on >2 Lanes" error also raised, overlook error |
0 Evaluate Results |
In addition, perform the following:
|
Only rerun specific bank of suspect lane | Escalate through your Regional Escalation Process |
| Liquid Level Sense Error | 1 Evaluate Result with septum Caps |
|
Only rerun specific bank of suspect lane without septum caps
result for the follow up run is 0 Evaluate Results without septum caps |
Escalate through your Regional Escalation process. |
|
Error in Full Fill Check
Description: Full fill check without amplification |
1 Evaluate Results |
|
Only rerun specific bank of suspect lane | Escalate through your Regional Escalation process. |
|
Normalizer on Color [color] Too Noisy
Description: Normalization noise error |
0 Evaluate Results |
|
On affected side, perform Full Qualification run to include QPM and EBP | Replace reader |
|
PCR Heater Warning
Description : PCR heater exceeded set tolerances |
0 Evaluate Results |
|
Only rerun specific bank of suspect lane | Replace heater/MUX board ONLY if same lane triggers error |
|
Step Loss on Axis Z; [how many] Steps
Description : Lost more than 15 steps on the Z axis |
0 Evaluate Results |
|
Rerun an EBP qualification run on both sides | Replace Gantry if test fails. |
|
Step Loss on Axis [X|Y]; [how many] Steps
Description : Lost more than 3 steps on the X/Y axis |
0 Evaluate Results |
|
Rerun an EBP qualification run on both sides | Escalate through your Regional Escalation process. |
|
Ct %CV > 5%; [percent]% ACROSS RUN
Description : Ct variation higher than expected across banks / readers |
0 Evaluate Results |
In addition, perform the following:
Evaluate resulting curves |
Rerun an EBP qualification run on both sides | Escalate through your Regional Escalation process. |
|
Ct %CV > 5% on Color [color]; [percent]% WITHIN Rack and Row [cartridge row]
Description : Ct variation higher than expected within same bank / reader |
0 Evaluate Results |
In addition, perform the following:
Evaluate resulting curves |
Only rerun specific rack | Escalate through your Regional Escalation process. |
|
Could not Pickup Tip
Description : Tip pickup failure on rack A/B |
0 Evaluate Results |
|
Rerun an EBP qualification run on both sides | Escalate through your Regional Escalation process. |
|
Heater Timeout Warning
Description : Lysis heater hardware issue |
0 Evaluate Results |
|
On affected side, perform an EBP qualification run | Replace GORT board and perform Full qualification run to include QPM and EBP |
|
PCR Detector Motor Failure Description : Reader detector motor issue |
0 Evaluate Results |
|
On affected side, perform Full Qualification run to include QPM and EBP | Escalate through your Regional Escalation process. |
| Errors Not Generated in Run Report | ||||
| Barcode Issues | ||||
| Cataloging Issue with Sample Tube Description: This condition is not generated in the run report and will occur during catalog | [0] Error/Run |
|
Run Catalog Script | Escalate through your Regional Escalation Process. |
| Cataloging issue with Snap ins Description: This condition is not generated in the run report and will occur during catalog | [0] Error/Run |
|
Run Catalog Script | Escalate through your Regional Escalation process. |
| Additional Issues | ||||
| Reader Saturation | N/A |
|
N/A | N/A |
|
Instrument Connection Loss During Run
Description: This condition is not generated in the run report and will generate an Alert |
0 Evaluate Results |
|
Perform Full Qualification run to include QPM and EBP | Escalate through your Regional Escalation process. |
|
Error in Empty Fill Check
Description: Empty fill check without amplification. This condition is not generated in the run report and will occur during catalog |
0 Evaluate Results |
|
On affected side, perform Full Qualification run to include QPM and EBP | Escalate through your Regional Escalation process. |
| Error in Results Processing | 0 Evaluate Results |
|
Repeat the qualification run | Escalate through your Regional Escalation process. |